Human IFNg (Interferon-gamma, IFN-g) ELISA Ready-SET-Go!
 
Print Page 
 
Adobe PDF  
Contents: Human IFNg (Interferon-gamma, IFN-g) ELISA Ready-SET-Go!
Catalog Number: 88-7316
Sensitivity: 4 pg/ml
Standard Curve Range: 4 - 500 pg/ml

Prices for This Product*
Cat. No. Size Price Qty Action
88-7316-22 2 plates (includes plates) $150
88-7316-76 20 plates (includes plates) $590
88-7316-77 20 plates $550
88-7316-86 10 plates (includes plates) $415
88-7316-88 10 plates $395
*International customers: Please contact your distributor for region specific pricing.

   
data image 1
Standard curve of human IFN-gamma ELISA Ready-SET-Go! Recombinant concentration in pg/ml.
Available Formats of This Product
Cat. No. Format Excite
(nm)
Emit
(nm)
Reported Applications
88-7316 Human IFNg (Interferon-gamma, IFN-g) ELISA Ready-SET-Go! N/A N/A ELISAset 

Questions? Please consult our answers to frequently asked questions at http://www.ebioscience.com/faq.


Description


This Human IFN-γ ELISA Ready-SET-Go! reagent set (with or without high-affinity binding microwell plates) contains the necessary reagents, buffers and diluents for performing quantitative enzyme linked immunosorbent assays (ELISA). This ELISA reagent set is specifically engineered for accurate and precise measurement of human IFN-γ protein levels from samples including serum, plasma, and supernatants from cell cultures.

Components

  1. Capture Antibody: Pre-titrated, purified antibody, clone NIB42
  2. Detection Antibody: Pre-titrated, biotin-conjugated antibody, clone 4S.B3
  3. Standard: Recombinant cytokine for generating standard curve and calibrating samples
  4. ELISA/ELISPOT Coating Buffer Powder: Reconstitute to 1L with dH20 and filter (0.22 uM).
  5. Assay Diluent: 5X concentrated
  6. Detection enzyme: pre-titrated Avidin-HRP
  7. Substrate Solution: Tetramethylbenzidine (TMB) Substrate Solution
  8. Certificate of Analysis: Lot-specific instructions for dilution of antibodies and standards
  9. 96 Well Plate: Corning Costar 9018 (included with product Cat. #’s ending in suffixes -22, -44, -76, -86)

References


Gupta AK, Holzgreve W, and Hahn W. 2005. Microparticle-free placentally derived soluble factors downmodulate the response of activated T cells. Hum Immunol. 66(9):977-84. (ELISA kit, PubMed)

Tao X, Li C, et al. 2007. Simultaneous detection of IFN-gamma and IL-4 in lesional tissues and whole unstimulated saliva from patients with oral lichen planus. J Oral Pathol Med. (ELISA kit, Epub)

Other Materials Needed

  • Buffers
    • Wash Buffer: 1 x PBS, 0.05% Tween-20 (or eBioscience ELISA Wash Buffer Powder, cat 00-0400)
    • Stop Solution: 1M H3PO4 or 2N H2SO4
  • Pipettes and pipettors
  • Refrigerator
  • 96-well plate (Corning Costar 9018 or NUNC Maxisorp flat-bottom)
    IMPORTANT NOTE: The use of ELISA plates which are not high affinity protein binding plates will result in suboptimal performance, e.g., low signal or inconsistent data. Do not use tissue culture plates or low protein absorption plates. Use only the Corning Costar 9018 or NUNC Maxisorp 96 well plates provided or suggested.
  • 96-well ELISA plate reader (microplate spectrophotometer)
  • ELISA plate washer

Stability

This ELISA set is guaranteed to perform as specified at least 12 months from date of receipt if stored and handled as instructed according to this datasheet and the Certificate of Analysis, which is included with the reagents.

Storage Instructions for Cytokine Standards

The frozen cytokine standard is already aliquoted at 20 µl per vial. Upon receipt, frozen cytokine standard should be immediately stored at -80°C; stable for at least 12 months. After thawing, quick-spin vial prior to opening. Do not re-aliquot into smaller fractions. These are single use vials. Use one time and discard. For dilution of the standard, please see instructions on the Certificate of Analysis and follow these as written.

Storage Instructions for Other Set Reagents

Store at 4°C.

IMPORTANT NOTE: Be certain that no sodium azide is present in the solutions used in this assay, as this inhibits HRP enzyme activity.

Time Requirements

  • 1 overnight incubation
  • 4˝-hour incubations
  • 1 hour washing and analyzing samples

Experimental Procedure


  1. Coat Corning Costar 9018 or NUNC Maxisorp 96 well ELISA plate with 100 µl/well of capture antibody in Coating Buffer* (dilute as noted on Certificate of Analysis, which is included with the reagent set). Seal the plate and incubate overnight at 4°C.
  2. Aspirate wells and wash 5 times with >250 µl/well Wash Buffer. Allowing time for soaking (~ 1 minute) during each wash step increases the effectiveness of the washes. Blot plate on absorbent paper to remove any residual buffer.
  3. Dilute 1 part 5X concentrated Assay Diluent with 4 parts DI water.* Block wells with 200 µl/well of 1X Assay Diluent. Incubate at room temperature for 1 hour.
    *Important: Do not include sodium azide in any buffers, as this will inactivate the HRP.
  4. Aspirate/wash as in step 2.
  5. Using 1X Assay Diluent, dilute standards as noted on the Certificate of Analysis (C of A). Add 100 µl/well of standard to the appropriate wells. Perform 2-fold serial dilutions of the top standards to make the standard curve. Add 100 µl/well of your samples to the appropriate wells. Cover or seal the plate and incubate at room temperature for 2 hours (or overnight at 4°C for maximal sensitivity).
  6. Aspirate/wash as in step 2. Repeat for a total of 5 washes.
  7. Add 100 µl/well of detection antibody diluted in 1X Assay Diluent (dilute as noted on C of A). Seal the plate and incubate at room temperature for 1 hour.
  8. Aspirate/wash as in step 2. Repeat for a total of 5 washes.
  9. Add 100 µl/well of Avidin-HRP* diluted in 1X Assay Diluent (dilute as noted on C of A). Seal the plate and incubate at room temperature for 30 minutes.
    *Important: Do not include sodium azide in any buffers, as this will inactivate the HRP.
  10. Aspirate and wash as in step 2. In this wash step, soak wells in Wash Buffer for 1 to 2 minutes prior to aspiration. Repeat for a total of 7 washes.
  11. Add 100 µl/well of Substrate Solution to each well. Incubate plate at room temperature for 15 minutes.
  12. Add 50 µl of Stop Solution to each well.
  13. Read plate at 450 nm. If wavelength subtraction is available, subtract the values of 570 nm from those of 450 nm and analyze data.

Ready-SET-Go Cytokine ELISA Set Buffers:

  • Assay Diluent (5X concentrate): Dilute 1/5 in DI water.
  • Substrate Solution: Ready to use (1X); 100 µl per well.
  • ELISA/ELISPOT Coating Buffer Powder: Reconstitute in 1L dH20; filter (0.22 uM).

Standard Calibration

The standard of the Ready-SET-Go! is calibrated against NIBSC standards:

Table 1: Table of Standard Calibration
Table of Standard Calibration
Cytokine ng of eB standard ng of NIBSC standard U of NIBSC standard NIBSC Lot #
hIL-2 1 1.1 14.6 86/564
hIL-4 1 2.2 22 88/656
hIL-5 1 2.2 22 90/586
hIL-6 1 1.7 170 89/548
hIL-10 1 0.8 4 93/722
hIL-12 1 0.8 8 95/544
hIFN-g 1 1.1 22 87/586
hTNF-a 1 0.9 36 87/650
mIL-2 1 3.1 310 93/566
mIL-4 1 3 30 91/656
mIL-6 1 8.5 850 93/730
mIFN-g* 1   4.5 Gg02-901-533
mTNF-a 1 1.7 340 88/532
* Mouse IFN-g is calibrated using NIH standard (Lot Gg02-901-533) and is measured in Units (U)

Table 2: ELISA Troubleshooting Guide
ELISA Troubleshooting Guide
Problem Possibility Solution
A. High Background 1. Improper and inefficient washing 1. Improve efficiency of washing. Fill plates completely, soak for 1 minute per wash, as directed
  2. Cross contamination from other specimens or positive control 2. Repeat ELISA, be careful when washing and pipetting
  3. Contaminated substrate 3. Substrate should be colorless
  4. Incorrect dilutions, e.g., conjugate concentration was too high 4. Repeat test using correct dilutions; check with the recommendations of the antibody manufacturer
B. No signal 1. Improper, low protein binding capacity plates were used 1. Repeat ELISA, using recommended high binding capacity plates
  2. Wrong substrate was used 2. Repeat ELISA, use the correct substrate
  3. Enzyme inhibitor present in buffers; e.g., sodium azide in the washing buffer and Assay Diluent inhibits peroxidase activity 3. Repeat ELISA, make sure your system contains no enzyme inhibitor
C. Very weak signal 1. Improper and inefficient washing 1. Make sure washing procedure is done correctly
  2. Incorrect dilutions of standard 2. Follow recommendations of standard handling exactly as written on the certificate of analysis
  3. Insufficient incubation time 3. Repeat ELISA, follow the protocol carefully for each step’s incubation time
  4. Incorrect storage of reagents 4. Store reagents in the correct temperature, avoid freeze and thaw, avoid using the “frost free” freezer
  5. Wrong filter in ELISA reader was used 5. Use the correct wavelength setting
  6. Wrong plate used 6. Use the recommended Corning Costar 9018 or NUNC Maxisorp flat bottom 96 well plates
D. Variation amongst replicates 1. Improper and inefficient washing 1. Make sure washing procedure is done correctly; see certificate of analysis
  2. Poor mixing of samples 2. Mix samples and reagents gently and equilibrate to proper temperature
  3. Plates not clean 3. Plates should be wiped on bottom before measuring absorbance
  4. Improper, low binding capacity plates were used 4. Use recommended high binding capacity plates
  5. Reagents have expired 5. Do not use if past expiration date